◈ Product Information
YMC-Pack PROTEIN-RP is a reversed-phase column utilizing a silica gel base. It contains a stationary phase, specifically designed for separation of proteins or peptides. Problems that are associated with conventional reversed-phase columns with short alkyl chain lengths are minimized.
Robust column lifetime and excellent recovery of hydrophobic proteins are typically possible on this phase.
► Point
• Improved recovery of proteins or peptides
• Improved durability when used with TFA solution
• Enabled elution of high molecular weight proteins
► Specification
- Particle size :5㎛
- Usable pH range :1.5-7.5
Improved recovery of proteins or peptides |
Sample |
Recovery (%) | ||
YMC-Pack |
C4 column, |
C4 column, | |
Ribonuclease A |
93 |
95 |
86 |
Cytochrome c |
94 |
89 |
95 |
Lysozyme |
98 |
93 |
76 |
Myoglobin |
97 |
85 |
93 |
BSA |
94 |
92 |
86 |
Ovalbumin |
90 |
73 |
88 |
Transferrin |
94 |
98 |
88 |
Insulin(bovine) |
97 |
73 |
92 |
Insulin B chain |
82 |
76 |
70 |
α-Mating factor |
93 |
82 |
81 |
Leu-Enkephalin |
92 |
84 |
91 |
Gly-Gly-Gly-Gly |
95 |
86 |
70 |
Recovery of various proteins and peptides is shown. YMC-Pack PROTEIN-RP gives greater recovery than competitive C4 columns.
Improved durability when used with TFA solution |
Flow conditions
• Eluent :water/TFA (100/0.1)
• Flow rate :1.0 mL/min
• Temperature :ambient
Measurement conditions
• Column :YMC-Pack PROTEIN-RP 250 x 4.6 mmI.D.
• Eluent :acetonitrile/water (40/60)
• Flow rate :1.0 mL/min
• Temperature :30ºC
• Detection :UV at 254 nm, 0.32AUFS
The graph shows the test result of the stability of the stationary phase with 0.1% aqueous TFA. Retention of diphenyl on C4 columns manufactured by other companies greatly decreases as time passes. This is caused by detachment of butyl groups from the packing material due to acid hydrolysis.
Retention of diphenyl on YMC-Pack PROTEIN-RP is shown to be stable after 500 hours of mobile phase flow.
◈ Ordering Information
YMC-Pack PROTEIN-RP Analytical Columns | ||
---|---|---|
Particle size(µm) |
Column size Inner diameter x length(mm) |
Product Number |
S-5 |
4.6 X 150 |
PR99S05-1546WT |
4.6 X 250 |
PR99S05-2546WT | |
YMC-Pack PROTEIN-RP Semi-preparative Columns | ||
Particle size(µm) |
Column size Inner diameter |
Product Number |
S-5 |
10 X 150 |
PR99S05-2510WT |
20 X150 |
PR99S05-1520WT |
|
20 X 250 |
PR99S05-2520WT | |
YMC-Pack PROTEIN-RP Guard Cartridge Columns | ||
Particle size(µm) |
Column size Inner diameter |
Product Number |
S-5 |
4.0 X 23 |
PR99S05-G304CC |
*When using it for the first time,please purchase the cartridge holder corresponding to inner diameter of column. |
YMC-Pack PROTEIN-RP Cartridge Holder | ||
---|---|---|
|
Product Number | |
Cartridge Holder (Inner diameter 4.0mm) |
XPCHW | |
YMC-Pack PROTEIN-RP Guard Columns | ||
Particle size(µm) |
Column size Inner diameter |
Product Number |
S-5 |
4.0 X 10 |
PR99S05-0104WFG |
10 X 30 |
PR99S05-0310WTG | |
20 X 50 |
PR99S05-0520WTG |